Abstract

A 60-year-old man with arrhythmogenic right ventricular cardiomyopathy was readmitted for the battery exchange of his implantable cardioverter-defibrillator (ICD). Since (i) he had been treated with a dual-coil shock lead (Sprint Fidelis, Medtronic) and (ii) pre-operative venography showed mild collateral flow to the left subclavian vein, a single-coil lead was additionally implanted. However, the single-coil defibrillation system was unable to terminate the induced ventricular fibrillation (VF), thus dual defibrillation shock pathways were created using the connection to the superior vena cava coil of the Fidelis lead. The combined connections of the two shock leads provided an appropriate margin of the defibrillation threshold.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.