Abstract
A label free, amperometric uric acid biosensor is described by immobilizing enzyme uricase through a self assembled monolayer (SAM) of 3-aminopropyltriethoxysilane (APTES) using a crosslinker, Bis[sulfosuccinimidyl]suberate (BS 3) on an indium-tin-oxide (ITO) coated glass plate. The biosensor (uricase/BS 3/APTES/ITO) was characterized by, scanning electron microscopy (SEM), atomic force microscopy (AFM) and electrochemical techniques. Chronoamperometric response was measured as a function of uric acid concentration in aqueous solution (pH 7.4). The biosensor shows a linear response over a concentration range of 0.05 to 0.58 mM with a sensitivity of 39.35 μA mM − 1 . The response time is 50 s reaching to a 95% steady state current value and about 90% of enzyme activity is retained for about 7 weeks. These results indicate an efficient binding of enzyme with the crosslinker over the surface of APTES modified ITO glass plates, which leads to an improved sensitivity and shelf life of the biosensor.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.