Abstract

We present a time-saving modification to the ultraviolet (UV) oxidation method for analyzing dissolved organic carbon (DOC) concentration, Δ14C, and δ13C measurements in seawater and standard materials. A low background (∼0.2 ± 0.2 μM) was reported for pre-irradiated Milli-Q (MQ) water that was used to dilute samples for DOC 14C analysis (Beaupré et al. 2007). We use MQ water without pre-irradiation (background ∼0.9 ± 0.2 μM) to dilute the sample. This method is suitable for small-volume, high-concentration samples (mass of sample DOC overwhelms mass of MQ water DOC). An acceptable precision of Δ14C measurements (5–9‰) is maintained. This revised method reduces the preparation time for diluted DOC Δ14C samples from 2 days to 1 day.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.