Abstract

The film of polyaniline/poly(acrylic acid) composite was prepared by electrochemical polymerization of aniline in the presence of poly(acrylic acid). An enzyme electrode was fabricated with the composite film by immobilizing glucose oxidase covalently on its surface. Amperometric glucose sensing was carried out with the enzyme electrode by monitoring O2 consumption accompanying enzymatic glucose oxidation. Prior to the glucose sensing, the electrochemical properties of the composite film were investigated with respect to cathodic O2 reduction. On the basis of the result, the glucose sensing was conducted at applied potential of −0.3V vs. SCE in a phosphate buffer solution of pH 7.0. The current response was found to increase linearly with an increase in glucose concentration up to 1.1mM. The sensitivity was 29.9±5.1μAmM−1cm−2 with the detection limit of 0.06±0.02mM (S/N=3). The response was not affected by such coexisting electroactive compounds as ascorbic acid and acetaminophen.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.