Abstract

This paper describes the results of a study undertaken to determine the sensitivity and reproducibility of a commercially produced alternating gradient force magnetometer. A noise base of 2×10−8 emu is achievable with an averaging time of 1 s. Reproducibility was found to be within 6% for most samples, but increases to 20% with low moment samples. The effect of the alternating field gradient on magnetic measurements has been fully investigated. The alternating field gradient is found to have significant effects on samples with a low fluctuation field and low coercivity. A low temperature system (4–300 K) is described and modifications to improve the sensitivity and base temperature are discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.