Abstract

Despite the use of ultraclean air, there are still cases of infection in total joint arthroplasty. One possible route by which bacteria may enter the wound is indirectly by contamination of instruments during skin preparation and draping. We found that bacterial air counts were 4.4 times higher during preparation and draping for hip or knee arthroplasty using an unscrubbed, ungowned leg holder than during the operation itself. With the leg holder scrubbed and gowned during preparation and draping, the air counts were reduced but were still 2.4 fold greater than intraoperatively. On some occasions, the air counts during preparation and draping exceeded the standards for ultraclean air irrespective of the attire of the leg holder. We recommend that the leg is held by a scrubbed and gowned member of the team. More importantly, we consider that instrument packs should be opened only after skin preparation and draping have been completed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.