Abstract
Preclinical single-photon emission computed tomography (SPECT) is an essential tool for studying the progression, response to treatment, and physiological changes in small animal models of human disease. The wide range of imaging applications is often limited by the static design of many preclinical SPECT systems. We have developed a prototype imaging system that replaces the standard static pinhole aperture with two sets of movable, keel-edged copper-tungsten blades configured as crossed (skewed) slits. These apertures can be positioned independently between the object and detector, producing a continuum of imaging configurations in which the axial and transaxial magnifications are not constrained to be equal. We incorporated a megapixel silicon double-sided strip detector to permit ultrahigh-resolution imaging. We describe the configuration of the adjustable slit aperture imaging system and discuss its application toward adaptive imaging, and reconstruction techniques using an accurate imaging forward model, a novel geometric calibration technique, and a GPU-based ultra-high-resolution reconstruction code.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Proceedings of SPIE--the International Society for Optical Engineering
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.