Abstract

A new triple-to-double coincidence ratio (TDCR) system has been established at CMI as an additional technique for primary activity standardisation. Details of the detectors and the electronics are given. Construction of the optical chamber was validated using measurement of a 3H standard—the efficiency achieved with Ultima Gold was 58%. Several methods of efficiency variation (coloured bands painted on vial, positioning and photomultipliers defocussing) are compared. Activity concentrations of 204Tl and 45Ca were determined with the TDCR method and with the efficiency tracer technique. Advantages of the TDCR method compared to the efficiency tracing method and agreement between results are shown.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.