Abstract

We have designed and constructed a quiet gradient set with restricted access for the combined purposes of evaluating the principles of active acoustic screening, recently introduced by Mansfield, Glover, and Bowtell, and for EPI studies of the head at 3.0 T. The design utilizes the return paths of the conductors in a closed arc loop arrangement to eliminate net Lorentz forces thereby attenuating acoustic noise especially at low frequency. This design should significantly reduce the dangers to patients of high noise levels, especially in high field magnetic resonance imaging systems.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.