Abstract

To analyze intraocular lens (IOL) surface abnormalities seen after folding using a forceps insertion technique. Mayo Clinic, Rochester, Minnesota, USA. Acrylic AcrySof MA60AC IOLs were examined using an Axio Imager microscope before and after they were folded using a forceps insertion technique. Differential interference contrast, brightfield reflected light, and darkfield reflected light imaging techniques were used as necessary. The effects of temperature, time, and ophthalmic viscosurgical device (OVD) on optic surface abnormalities after folding were studied. All 17 IOLs examined had smooth, defect-free optic surfaces before folding. After folding, anterior optic surface depressions were observed in all IOLs; the depressions corresponded to the contact area of the titanium insertion forceps. Surface depressions were present up to 72 hours after folding, were more pronounced when an insertion forceps with a high degree of wear was used, and were greater when the IOL was warmed to 98 degrees F before folding. Coating the IOL surface with OVD before grasping it with the insertion forceps prevented formation of depressions. The anterior optic surface of the acrylic IOL was vulnerable to forceps-induced surface depressions. Surface abnormalities were prevented by coating the anterior optic surface with OVD before grasping it with a metal insertion forceps.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.