Abstract

We have measured the differences between the Kelvin thermodynamic temperature and the temperature of the International Temperature Scale of 1990 on nine isotherms between the triple point of mercury and 380 K, by means of a primary acoustic thermometer. For the present measurements the standard uncertainty of (T − T90) ranges from 0.9 mK at 234 K to 1.7 mK at 380 K. The experimental method is based on the measurement of the acoustic resonance frequencies of an argon-filled spherical cavity and the microwave resonance frequencies of the same cavity when evacuated. The present results agree within the remarkably small combined uncertainties with both NIST acoustic thermometry ([1] Moldover M R et al 1999 J. Res. Natl Inst. Stand. Technol. 104 11–46; [2] Strouse G F et al 2002 Progress in primary acoustic thermometry at NIST: 273 K to 505 K 8th Temperature Symp. (Chicago, 21–24 October 2002)) and UCL acoustic thermometry ([3] Ewing M B and Trusler J P M 2000 J. Chem. Thermodyn. 32 1229–55) in the overlapping temperature range.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.