Abstract

There are few studies on the effect of acetabular cup design on cement penetration. We evaluated the effects of an acetabular flange on cement pressurization and cement penetration in 12 cadavers. Flanged or unflanged cups were implanted in paired human acetabula with simulated intraosseous bleeding pressure but without cement pressurization before insertion of the cup. Three pressure transducers were used to record intra-acetabular peak and average pressures during cup insertion. Following implantation, the whole specimens were AP-radiographed and standardized sections through the acetabula were microradiographed to evaluate cement penetration. Flanged cups produced greater intra-acetabular peak pressures than unflanged cups, but did not increase the average intra-acetabular pressure. Cement penetration did not differ significantly between the two groups. Our findings do not support the use of flanged cups as the sole means of cement pressurization in the acetabulum.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.