Abstract

We present a description of a polarizer rotating-polarizer compensator spectroscopic ellipsometer called PRPCSE. This configuration is particularly adapted to characterize transparent film/substrate arrangement. We developed a procedure for the calibration with a single configuration of the optical elements (i.e. the same configuration for the calibration and the measurement). This method depends neither on the sample used nor on the gain of the ellipsometer, and the exact knowledge of the compensator phase shift value is not necessary. As for the measurements, this method allows to determine the tan Ψ and tan Δ parameters of the sample and is adapted to minimize systematic errors. It presents a special advantage when Δ is close to zero because the experimental errors are proportional to cos 2 Δ compared to a conventional rotating polarizer ellipsometer where the experimental errors are proportional to sin −1 Δ. We show the potentialities of this ellipsometer for the measurements of transparent substrate with a sample of CuCl nanocrystals in NaCl matrix. A good precision of the measured parameters was obtained.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.