Abstract

Abstract The YSI model 2300 Stat Plus analyzer accurately measures glucose in aqueous solutions with an overall standard error of prediction (SEP) of 0.28 mM and a mean percent error (MPE) of 1.79 %. The absolute and relative errors for glucose measurements are concentration dependent. Initial measurements demonstrate the need to periodically monitor the accuarcy of this analyzer to avoid systematic errors. Lactate is measured with an overall SEP of 0.18 mM and MPE of 5.10 %. Analysis of the residuals indicates a slight negative bias of -0.07 mM for lactate.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.