Abstract

To evaluate the sensitivity of the Securetec DrugWipe 6 S Ketamine device in detecting salivary ketamine, we administered three fixed, increasing sub-anaesthetic doses of intravenous ketamine solution to 20 participants using an open-label, within subjects’ protocol. Results indicated, at a cut-off level of 5 ng/mL for ketamine in oral fluid, the Securetec DrugWipe 6 S Ketamine device indicated a uniformly positive result for the presence of ketamine during the acute dosing phase where corresponding mean whole blood ketamine concentrations were 0.38 mg/L (dose 1), 0.48 mg/L (dose 2) and 0.82 mg/L (dose 3), respectively. At 2-hours post treatment, the positive detection rate was 85% where corresponding mean whole blood concentrations of ketamine were 0.21 mg/L. Device detection rate reduced to 40% at 4-hours post-treatment.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.