Abstract

This study assessed the accuracy of an electronic hinge axis tracing device (Cadiax compact) in measuring the horizontal condylar inclination (HCI) and the Bennett angle. The hinge axis movements were simulated with an articulator to which the hinge axis tracer was attached. On the articulator the pre-set HCI values were 20 degrees, 40 degrees and 60 degrees, the pre-set Bennett angles were 0 degrees, 5 degrees, 10 degrees, 15 degrees, 20 degrees. The maximum measuring errors ranged from 0 to 3.4 degrees (in average 1.2 degrees ) and yielded statistically significant differences between articulator setting and Cadiax compact measurement (P < 0.05) except the HCI of 40 degrees and the Bennett angle of 0 degrees. Because of the small range of the maximum measuring error the Cadiax compact represented reasonable accuracy for clinical application in anterior guidance restorations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.