Abstract
Accuracy of total knee arthroplasty (TKA) implant placement and overall limb are important goals of TKA technique. The accuracy and ease of use of an accelerometer-based hand-held navigation system for tibial resection during TKA was examined in 90 patients. Preoperative goals for sagittal alignment, navigation system assembly time, resection time, and tourniquet time were evaluated. Coronal and sagittal alignment was measured postoperatively. The average coronal tibial component alignment was 0.43° valgus; 6.7% of patients had tibial coronal alignment outside of ±3° varus/valgus. The difference between the intraoperative goal and radiographically measured posterior tibial slope was 0.5°. The average time to completion of the tibial cut was 4.6 minutes. The accelerometer-based hand-held navigation system was accurate for tibial coronal and sagittal alignment during TKA, with no additional surgical time compared with conventional instrumentation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.