Abstract

To evaluate the accuracy and variability of right ventricular (RV) volumes and mass using dual-source computed tomography (DSCT) and the influence of slice orientation in comparison to cardiac magnetic resonance imaging (CMR). In 33 patients undergoing cardiac DSCT and CMR, RV parameters were calculated using the short-axis (DSCT, CMR) and axial orientation (DSCT). Intra- and interobserver variability were assessed by Bland-Altman analysis. Short-axis orientation: RV parameters of the two techniques were not statistically different. Axial orientation: RV volumes and mass were significantly overestimated compared with short-axis parameters whereas EF was similar. The short-axis approach resulted in low variability, although the axial orientation had the least amount of intra- and interobserver variability. RV parameters can be more accurately assessed by DSCT compared with CMR using short-axis slice orientation. RV volumes and mass are significantly higher using axial compared with short-axis slices, whereas EF is unaffected. RV parameters derived from both approaches yield high reproducibility.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.