Abstract

The orientation of the acetabular component after total hip arthroplasty is expressed with inclination angle and anteversion angle (VA). We studied the differences in accuracy and precision between VA measured on AP radiographs of the pelvis and those measured on cross-table lateral radiographs when compared to the gold standard measurements on CT using a supine functional pelvic plane as reference. VA was measured for 66 hips at one week after surgery on an AP radiograph of the pelvis (VAP), a cross-table lateral radiograph of the hip (VCL), and CT of the pelvis. Mean (± SD) of error was 2.8° ± 4.1° for VCL, and -0.57° ± 3.1° for VAP. VAP showed a higher accuracy and precision than VCL. For measurements of VA, we recommend that they be made on an AP radiograph of the pelvis rather than a cross-table lateral radiograph of the hip.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.