Abstract
Eight patients undergoing acetate-free biofiltration (AFB) suffered aluminum intoxication. The source of this outbreak was parenteral exposition to high concentrations of aluminum in sodium bicarbonate solutions. The manufacturer of bicarbonate solutions used in AFB was substituted in May 1994 and the solutions were stored in glass containers. At the peak of intoxication (July 1994) serum aluminum determination revealed an average value of 147.3 ± 21 µg/l. Aluminum levels in bicarbonate solutions were 400 µg/l. Serum ferritin rose from 307.4 ± 161 to 735.6 ± 206 ng/ml, whereas MCV decreased significantly from 98.4 ± 9 to 90.1 ± 10 fl. No significant changes were found in hemoglobin, neither in plasma iron, nor in iron transferrin saturation. The doses of recombinant human erythropoietin showed a considerable increase. The replacement solutions were changed and a new solution, stored in plastic containers and with aluminum levels lower than 10 µg/l, was used. The biochemical parameters were normalized. This outbreak demonstrates the need for a stringent control of aluminum-containing replacement fluids.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.