Abstract

The National Metrology Institute of South Africa (NMISA) has implemented a system to measure the transverse sensitivity of vibration transducers. As a mechanical device, the principle sensing axis of an accelerometer is not 100 % perpendicular to the mounting axis. This gives rise to the effect that the accelerometer will produce an electrical output even when a mechanical input perpendicular to the principle measurement axis is applied. The quantification of this "defect" parameter is of importance when high accuracy acceleration measurements are performed using accelerometers. This paper gives a brief overview of the system developed by the NMISA to measure the transverse sensitivity of vibration transducers. The paper then explores the validation of the system along with the uncertainty of measurement associated with the calibration system.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.