Abstract
The radioactivity of 60Co and 134Cs sources were measured using a 4 πβ–4 πγ detector configuration with a well-type NaI(Tl) crystal and a sandwich type 4 πβ detector composed of two sheets of NE102A plastic scintillator coupled to a slender photomultiplier tube. The β-detector was inserted into the well of the γ-detector. Since counting efficiencies in both the β- and γ-channels can be kept high, excellent counting statistics were attainable even when weak sources were measured. This configuration can also be used for radioactivity measurements based upon the direct integral counting of 4 πβ+4 πγ logic sum signals, and two independent results can be obtained for every measurement. This technique is especially useful for the standardization of complex decaying nuclides.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.