We establish that clamping effects, which limit accurate determination of piezoelectric responses in bulk materials and films using piezoelectric force microscopy (PFM), are not present when measuring discrete nanostructures with radii less than five times the tip radius. This conclusion is established by comparing the piezoelectric response in ZnO rods using two electrode configurations: one with the conducting atomic force microscopy tip acting as the top electrode and the other using a uniform metal top electrode. The distributions of piezoelectric coefficients measured with these two types of electrode configurations are the same. Hence, clamping issues do not play a role in the piezoelectric property measurement of nanomaterials using PFM. The role of conduction electrons on the piezoelectric measurement in both cases is also discussed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.