Abstract

A sensitive GC/MS method for the determination of amisulpride in whole blood was developed, optimized and validated. Sample preparation included solid-phase extraction using HF Bond Elut C18 cartridges and further derivatization with heptafluorobutyric anhydride (HFBA). The limits of detection and quantification were 3.00 and 10.0 μg/L, respectively. The calibration curves were linear up to 1000 μg/L (R(2)≥0.991). Absolute recovery ranged from 94.2 to 101%. Accuracy was found to be between -8.7 and 1.9% and imprecision was less than 10.0%. The developed method covers the generally accepted therapeutic range but it can also cover levels above them. This makes our method suitable for the determination of amisulpride not only for clinical purposes on psychiatric patients, but also during the investigation of forensic cases where amisulpride is involved.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.