Abstract
An enzyme-labeled immunometric assay has been developed for measuring digoxin concentrations in serum or plasma. Unitized, compartmentalized reagents are used with an automated sample-processing instrument. The enzyme activity of the processed sample, which is directly proportional to the digoxin concentration, is measured by using a reagent strip and the Ames Seralyzer reflectance photometer. The test takes less than 15 min, and digoxin concentrations are calculated from a two-point calibration line stored in the instrument. Within-run CVs for controls at four concentrations ranged from 2.3% to 3.8%; between-run CVs were from 1.5% to 2.6%. Results obtained with clinical serum samples correlated well (r greater than 0.96) with those obtained by fluorescent polarization immunoassay (Abbott TDx) and RIA (Clinical Assays and NML). This rapid and convenient method for monitoring digoxin concentrations in serum or plasma is particularly well suited for decentralized sites such as emergency rooms, urgent-care centers, and physicians' offices.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.