The circuit described has been installed in a medical linear accelerator to operate the high voltage circuit interlock whenever arcing occurs in the high voltage pulse forming network (PFN) or should free-running occur in the trigger pulse generator for the main thyratron. The PFN charging pulses are monitored by a 10 Omega resistor placed in series with the high voltage power supply capacitor and fed directly into the circuit.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.