Abstract

Theoretical expressions have been derived for the contributions to ionisation density in cavity chambers from secondary electrons arising from photon absorption in wall, build-up cap and surrounding medium. The method allows the derivation of Clambda and displacement factors, which are evaluated for the Farmer graphite chamber for a range of photo beams. Differences between Clambda values calculated for air-walled and water-walled chambers are found to be consistent with those determined by Nahum and Greening (1978). The currently recommended displacement. factor for the Farmer chamber with cap, however, is found to be in error by about 0.8%. On the basis of a comparison between theoretically predicted air doses in cavities and the experimental results of Fregene (1975) it is proposed that no further correction to Clambda is required for the chamber 'effective point of measurement'.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.