Abstract

Ultrasound image uniformity is an important parameter for quality assurance in diagnostic ultrasounds, but it is usually assessed by a qualitative judgement of technicians so its estimation is rough and subjective. In this work a novel method is developed to give an objective measurement of the Bmode image uniformity over the whole field of view or some of its part: the Texture Distribution Analysis Method (TDAM) is based on a segmentation of the Region of Interest, depending on some texture features calculated from co-occurrence matrices. Results on a set of 10 test images with different non-uniformities (Uniformity Image Test Set or UITS) show a good sensitivity and agreement of TDAM with the mean judgment by 5 human observers (TUV): TDAM and TUV uniformity values are coherent for the whole UITS, nevertheless a high uncertainty in uniformity values has been observed (up to 28 percent). Preliminary results look encouraging and more efforts are worth to refine the method.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.