Abstract

Transcutaneous energy transmission systems for the artificial heart with synchronous rectifiers were developed. The input voltage is 24 V and the output voltage is 9 to 18 V. This paper shows the efficiency of the system with synchronous rectifier using a digital PLL technique achieved 92% DC-DC efficiency at 5 mm coil distance. Using the ferrite core and a digital PLL technique, inductor and capacitor were eliminated from the rectifier circuit. And the circuit was mounted to the back of the secondary coil. And we developed the control systems for stable output voltage and apply this result to other contact power supply system.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.