Abstract

This work presents an investigation about the type 3 non-uniqueness of the International Temperature Scale of 1990 (ITS-90) between the aluminium (660°C) and silver (962°C) freezing points, carried out at the Italian National Institute of Metrology (INRiM) within the framework of the EMRP—JRP SIB10 NOTED (Novel Techniques for Traceable Temperature Dissemination). For this purpose six High Temperature Standard Platinum Resistance Thermometers (HTSPRTs) from different manufacturers were calibrated at the ITS-90 fixed points, then intercompared in a sodium Gas Controlled Heat Pipe (GCHP) at 25K intervals between the aforementioned fixed points; a temperature uniformity and stability within ±0.5mK was achieved in the whole range between 660°C and 962°C. This paper presents the results of type 3 non-uniqueness analysis of these test HTSPRTs, together with problems deriving from their stability, which is the largest source of uncertainty in this work.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.