Abstract

A prototype clinical brain single-photon emission computed tomography (SPECT) insert has been designed for use in simultaneous SPECT/MRI. The system utilizes novel slit-slat collimators which, like pinhole collimators, suffers from parallax errors due to the large incident angle of photons. A statistical algorithm has been developed to determine the depth-of-interaction (DOI) with a view to improving image performance. The importance of DOI correction was demonstrated using Monte Carlo simulation. This simulation also indicated that four DOI layers ( <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$3 \times 1.5\, \textrm {mm} + 3.5\, \textrm {mm}$ </tex-math></inline-formula> ) may be sufficient. The improvement in event localization was demonstrated on a single detector before implementing the algorithm on the full clinical prototype where some limitations in event localization in layers close to the readout plane were observed. Nevertheless, DOI enabled the rejection of poorly localized events with improved resolution in reconstructed line sources.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.