Abstract

A device was designed to provide a "square" pulse of blood flow into the arterial system. Pulses were injected into the carotid artery of the rabbit during transient cardiac arrest. Analysis of pressure response curves generated by the flow provides information as to the state of the arterial tree. With certain assumptions it is possible to estimate from these curves lumped values of peripheral resistance, critical closing pressure, and arterial compliance. In a series of 12 rabbits the mean value of peripheral resistance was found to be 0.21 +/- 0.7 mmHg-ml-1-min and critical closing pressure was estimated to be 23.6 +/- 3.8 mmHg. This method gives two possible values for arterial compliance 0.036 +/- 0.010 and 0.055 +/- 0.010 ml-mm-1 based, respectively, on the rise and decay curves of the pressure response. The theory and limitations of the method are discussed. The use of the method is illustrated in following the response to increased PCO2 and hemorrhage.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.