Abstract

We report a new configuration of a reflection-type confocal scanning optical microscope system for measuring the refractive index profile of an optical waveguide. Several improvements on the earlier design are proposed; a light emitting diode at 650nm wavelength instead of a laser diode or He–Ne laser is used as a light source for better index precision, and a simple longitudinal linear scanning and a curve fitting techniques are adapted instead of a servo control for maintaining an optical confocal arrangement. We have obtained spatial resolution of 800nm and an index precision of 2×10−4. To verify the system’s capability, the refractive index profiles of a conventional multimode fiber and a home-made four-mode fiber were examined with our proposed measurement method.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.