Abstract

An improvement in the uniformity of venetian-blind photomultiplier tubes has been achieved by reducing the voltage difference between the first and second dynodes. The method has been applied to a gas proportional scintillation counter (GPSC) instrumented with a venetian-blind photomultiplier (PMT). When exposed to a 20-mm collimated 5.9-keV X-ray beam, an overall improvement in energy resolution for the GPSC/PMT combination from 20% to 11.5% was achieved. An alternative method that reduces the photocathode-to-first-dynode voltage was less effective and resulted in a severe degradation of detector energy resolution.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.