Abstract
A simplified peak fitting technique for the analysis of the overlapped pulse-height spectra of 134Cs and 137Cs γ-rays obtained with a NaI(Tl) scintillation spectrometer was studied. In this analysis, nearly an upper half of 662keV 137Cs peak data was employed for fitting of the Gaussian peak using the least squares algorithm. Consistent results were obtained as compared with the reference value of test samples mixed with 134Cs and 137Cs standard solutions.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.