The diminution of exposure rate from radium gamma radiation by filtration through platinum can be described very accurately by gamma 0/(1 + mu0d) for wall thickness d from 0.2 mm to 4.0 mm (gamma 0 = 6.356 x 10(-5) A m2/kg2 or 8.869 R cm2 h-1 mg-1; mu0 = 1.501 cm-1). This function fits generally accepted data to within experimental error, to an accuracy not achieved by any single exponential factor. The expression is derived theoretically from a simple approximation to the spectral distribution of radium gamma radiation. With this type of transmission function, the expression for the exposure rate around linear radium sources in platinum tubes (generalized Sievert integral) is readily integrated in terms of elementary functions. Used in place of the Sievert integral in calculations of dose distributions, this new fomulation makes possible straightforward numerical evaluation and avoids the storage of tabulated values in computer programmes.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.