Abstract

A new analytical procedure for fast and simultaneous determination of sulfamethoxazole (SMX), trimethoprim (TMP) and phenazopyridine (FZP) has been developed. The method is based on batch injection analysis with multiple pulse amperometric detection (BIA-MPA) which offers several desirable features, such as simplicity, reduced costs, robustness, good precision (RSD <3.5%, n=15), high analytical frequency (up to 70injectionsh−1) and generation of small volume of waste by analysis. The BIA-MPA procedure provides linear range from 4 to 320, from 2 to 40 and from 1 to 40mgL−1 and limits of detection of 0.20, 0.15 and 0.05mgL−1 for SMX, TMP and FZP, respectively. The results obtained with the proposed method were compared to those obtained by HPLC and similar results were obtained (95% confidence level).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.