Abstract

In this study, the initial report on determining cabergoline via nanostructure-adjusted ionic liquid carbon paste electrode with aqueous solutions is described. For this purpose, an original adjusted carbon paste electrode that uses SiO2@Fe3O4/GO nanocomposite and 1-methyl-3-butylimidazolium bromide as a binder (SiO2@Fe3O4/GO/CPILE) was designed. Cabergoline oxidation peak at SiO2@Fe3O4/GO/CPILE surface was 500 mV that was approximately 200 mV less compared to the oxidation potential pertaining to the CPE surface subjected to a similar state. Moreover, there was an increase in the peak current of approximately 3.0 times greater at SiO2@Fe3O4/GO/CPILE surface in comparison to that of the CPE. The relevant linear response range and detection limit were determined as 0.07–500.0 and 0.01 μM, correspondingly. The adjusted electrode was successful in determining cabergoline within real specimens entailing adequate results.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.