Abstract

We have developed a novel, semiautomated carotid intima-media thickness (CIMT) border detection program (AUTO) and evaluated its measurement reproducibility and accuracy. Images from 6 carotid segments were acquired in 50 subjects, for a total of 300 segments. Mean and maximum CIMT values were measured blindly at a reference (REF) lab and in duplicate by experienced (EXP) and novice (NOV) readers using manual (MAN) and AUTO methods. Coefficients of variation for AUTO measurements of mean (3.2%) and maximum (4.1%) CIMT were low, and the AUTO method improved the NOV reader's reproducibility. Compared with the REF lab, mean (0.012 +/- 0.006 mm) and maximum (0.144 +/- 0.006 mm) CIMT biases were small and equivalent to those of the REF lab ( P < .001). The AUTO method shortened reading times by 35% to 46% ( P < .001). We conclude that our novel AUTO CIMT measurement program improved reproducibility and was accurate. Compared with MAN tracing, the AUTO method agreed better with the REF lab and decreased reading time.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.