Abstract

Positron emission tomography (PET) image quality in both clinical and preclinical environments highly depends on an accurate knowledge of the detector hardware to correct for image quality degrading effects like gain, temperature, and photon detection efficiency variations of the individual crystals. In conventional PET systems some of these variations are typically corrected using a dedicated calibration scan in which the scanner performance for a well-known activity source is measured. We propose an alternative method for estimating the relative sensitivity of each detector pixel using the coincidences as well as the singles emission data of each PET scan. The overall idea is to compare the total sum of all measured single photons before coincidence processing in each crystal with a steadily low-frequent distribution that can normally be expected. Both the estimated activity and the estimated detector sensitivity are simultaneously improved by using an extended iterative reconstruction scheme. This way we ensure the use of an optimal calibration correction (with the exception of a global factor) for each data set, even if the scanner performance has changed between two scans. Data measured with a preclinical PET scanner (HYPERIon-I) which uses analog silicon photomultipliers in combination with a custom-made ASIC shows a significant increase of image quality and homogeneity using the proposed method.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.