Abstract
For the ultimate sensitivity in trace radiochemical analysis, the radiation detector must have high efficiency and low background. A low-background gamma-ray spectrometer in regular use at NIST for over twenty years is being supplemented by a new system, improved in several ways. The new detector is much larger, a shield of iron reduces cosmic neutron background compared with lead, large plastic scintillators reduce the muon continuum background, and a digital data acquisition system gives new opportunities for optimization.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.