Abstract
A rolling ball viscometer has been developed for use at high pressures up to 1 GPa. It is proved that the calibration coefficient of the viscometer depends on its inclination angle and this dependence varies with the diameter ratio of the ball and tube. Improvement is made in order to allow simultaneous measurement of the density by assembling a potentiometric displacement meter. The precision of the density measurement is estimated to be ±0.22%. The viscosity data of di-(2-ethyl-hexyl) sebacate (Univis P12) is presented in the range of pressures up to 800 MPa at temperatures from 6 to 60°C. The accuracy of the viscosity measurement is estimated to be ±0.95%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Physica B+C
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.