Abstract

We describe an autocollimating optical angle sensor with a dynamic range of 9 mrad and nrad/\documentclass[12pt]{minimal}\begin{document}$\sqrt{\textrm {Hz}}$\end{document} Hz sensitivity at frequencies from 5 mHz to 3 kHz. This work improves the standard multi-slit autocollimator design by adding two optical components, a reference mirror and a condensing lens. This autocollimator makes a differential measurement between a reference mirror and a target mirror, suppressing common-mode noise sources. The condensing lens reduces optical aberrations, increases intensity, and improves image quality. To further improve the stability of the device at low frequencies the body of the autocollimator is designed to reduce temperature variations and their effects. A new data processing technique was developed in order to suppress the effects of imperfections in the CCD.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.