Abstract
A real-time ultrasound time-domain correlation (UTDC) blood flowmeter has been developed. Real-time performance has been achieved through the implementation of a custom-designed high-speed residue-number system (RNS) hardware correlator. The flowmeter is interfaced to a commercial ultrasound imager and can produce one-dimensional velocity versus range graphs at a rate of three per second. It has been validated in a blood flow phantom under a variety of conditions along with in vivo measurements in the human carotid artery. The theory of the time-domain correlation technique, design and implementation of flowmeter hardware, and the important correlation parameters which affect the performance of the flowmeter are described.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.