Abstract

This study aimed to demonstrate potential measures to discern interference and establish an algorithm for identifying interference in an hs-cTnT assay with a clinical discrepancy. Laboratory error and disease factors were first excluded. Then, a serial dilution test, polyethylene glycol (PEG) precipitation, and heterophile antibody blocking reagents (HBRs) were performed to confirm the interference substance. In addition, high-speed centrifugation and western blot were tried. The sample diluted linearly (R2 = 0.9966). However, there was a dramatic reduction in the concentration of hs-cTnT concentration to a normal level after both PEG precipitation and HBR1 treatment. The results were inconclusive in high-speed centrifugation and western blot assay. This study elucidated a heterophile antibody interference in measurement of hs-cTnT alone on a Roche analyzer.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.