Abstract

A gas chromatographic method using nitrogen-phosphorus detection was developed to quantitate clozapine in plasma or serum. Methyl clonazepam was used as an internal standard. Sample preparation included a single-step extraction with ethyl acetate, which was injected directly onto a wide-bore capillary column. Within-run and total precision, measured as percent coefficient of variation, were determined at low, therapeutic, and high clozapine plasma concentrations. The within-run precision for the low, therapeutic, and high clozapine plasma samples was 5.2, 2.7, and 2.4%, respectively. The total precision for the low, therapeutic, and high clozapine plasma samples was 10.0, 2.6, and 2.0%, respectively. Analytical accuracy was evaluated by comparing quantitative results with those obtained from a reference laboratory. Those samples containing therapeutic or high concentrations agreed within 3%; the sample containing a subtherapeutic concentration differed by 11.9%. The limit of quantitation was determined to be 35 ng/mL, and the upper limit of linearity was 3000 ng/mL. No significant interferences were detected after testing more than two dozen drugs and metabolites.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.