Abstract

Cross-linked liners were introduced with the promise that they would substantially reduce polyethylene wear. In 1999, our institution initiated a prospective study to compare the outcome of total hip arthroplasty patients who were randomized to non–cross-linked Enduron liners with that of total hip arthroplasty patients who were randomized to Marathon polyethylene liners that had been cross-linked with 5 Mrad (50 kGy) of gamma-irradiation and heat-treated to eliminate free radicals. At a mean follow-up of 5.7 years, the clinical outcomes among the Marathon and Enduron liners were similar. However, the mean wear rate was 0.01 ± 0.07 mm/y for the Marathon group, which represents a 95% reduction compared with the mean wear rate of 0.19 ± 0.12 mm/y for the Enduron group. In addition, the incidence of osteolysis was lower in the Marathon group.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.