Abstract

Hyperpolarized carbon-13 magnetic resonance spectroscopy (MRS) is a powerful tool to explore tissue metabolic state, by permitting the study of intermediary metabolism of biomolecules in vivo. However, a number of technological problems still limit this technology and need innovative solutions. In particular, the low molar concentration of derivate metabolites give rise to low signal-to-noise ratio (SNR), which makes the design and development of dedicated radiofrequency (RF) coils a fundamental task. In this article, the authors describe the simulation and the design of a RF coils configuration for MR experiments in mice, constituted by a 1 H whole body volume RF coil for imaging and a 13 C single circular loop surface RF coil for performing 13 C acquisitions. After the building, the RF system was employed in an in vivo experiment in a mouse injected with hyperpolarized [1-13 C]pyruvate by using a 3 T clinical MR scanner. SCANNING 38:710-719, 2016. © 2016 Wiley Periodicals, Inc.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.