Abstract
A pulsatile pump driven by a coil spring, which was designed and constructed by us, is described in this report. It consists of two main parts, a disposable blood chamber and a driving section. The blood chamber has two leaflet valves and a piston, which is covered with two bellofram rolling diaphragms and moves into the housing to draw in and eject the blood. The driving section consists of three cams, an electric motor and a coil spring. The ejection force is wholly produced by the compressed coil spring and is transmitted to the piston in the blood chamber by a rod. This pump allows the ejection pressure, the beat rates, and the stroke volume all to be changed independently. The performance of the pump was tested by using a circulation model where the beat rate was adjusted from 30 to 250 bpm. The output subsequently increased from 0.8 l/min to 5.7 l/min and the stroke volume, from 20.4 ml to 36.7 ml. This new pump has been used for clinical cardiopulmonary bypasses in 24 patients of open heart surgery and the pressure traces during perfusion resembled those of the patients' own hearts.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.