Abstract

A proximal positive end-expiratory pressure (PEEP) and continuous positive airway pressure (CPAP) system was constructed by the simple addition of a venturi T-piece proximal to the exhalation limb of a breathing circuit. The level of PEEP or CPAP was determined by the amount of flow powering the venturi. This system can provide positive pressure in excess of 40 cm H2O without the need for check valves, dump valves or additional nebuliers and flowmeters.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.